Alternative fiducial markers for Vero real-time tumor tracking radiotherapy: A phantom study

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

The objective of this study was to investigate the feasibility of potential fiducial markers consisting of various materials in a Vero real-time tumor-tracking (RTTT) system. In order to determine the applicability of fiducial markers for the Vero RTTT system, we tested various markers consisting of 8 kinds of material (titanium, stainless steel, high-carbon steel, pure steel, copper, silver, tantalum, and gold) with various diameters ranging from 0.3 mm to 1.6 mm and a length of 5 mm. Additionally, a commercial gold coil marker (Visicoil™, IBA dosimetry, Schwarzenbruck, Germany) of diameter 0.5 mm and length 1 cm was included for evaluation. The radiologic visibility on kV fluoroscopy/kV CT scan images of the fiducial markers was evaluated. The detectability on the RTTT system was tested using a two-dimensional moving phantom (Brainlab AG, Feldkirchen, Germany), producing sinusoidal motion. The target center’s accuracy was evaluated by calculating the deviation of the position of a metal sphere from the center on the dose profile. Dose profiles were measured using Gafchromic EBT2 films (International Specialty Products, NJ, USA). All markers were visible on kV fluoroscopy/kV CT while markers with atomic number ≥ 25.7 were detectable on the Vero RTTT system. All the detected markers showed excellent geometric accuracy.

Original languageEnglish
Pages (from-to)1700-1704
Number of pages5
JournalJournal of the Korean Physical Society
Volume69
Issue number11
DOIs
StatePublished - 1 Dec 2016

Keywords

  • Fiducial marker
  • Four-dimensional image-guided radiotherapy
  • Real-time tumor tracking
  • Respiratory motion
  • Vero

Fingerprint

Dive into the research topics of 'Alternative fiducial markers for Vero real-time tumor tracking radiotherapy: A phantom study'. Together they form a unique fingerprint.

Cite this