Autonomous calibration method of the reference flat surface of an interferometer without using a standard flat surface

Ikumatsu Fujimoto, Toshiyuki Takatsuji, Kunitoshi Nishimura, Min Young Kim

Research output: Contribution to journalArticlepeer-review

16 Scopus citations

Abstract

An autonomous method for calibrating the reference flat surface of an interferometer is proposed with the uncertainty analysis. The method consists of three phases; the first step is multiple rotating shifts of a specimen, the second is a linear shift, and the last is multiple rotating shifts again. The profile of the reference flat surface is basically determined by the linear shift. The linear shift errors that occurred during the linear shift are identified by the rotating shifts. The rotating shift errors caused by the rotating shifts can be compensated and the residual uncertainty can be reduced in proportion to the square root of the number of rotating shifts per one revolution. Finally, the uncertainty analysis is carried out in detail.

Original languageEnglish
Pages (from-to)4754-4767
Number of pages14
JournalApplied Optics
Volume51
Issue number20
DOIs
StatePublished - 10 Jul 2012

Fingerprint

Dive into the research topics of 'Autonomous calibration method of the reference flat surface of an interferometer without using a standard flat surface'. Together they form a unique fingerprint.

Cite this