Development of liquid scintillator system for proton flux monitoring

J. H. So, H. J. Kim, Heedong Kang, H. Park, S. Ryu, S. W. Jung, Sihong Doh, Sunghwan Kim, Kyeryung Kim

Research output: Contribution to journalArticlepeer-review

13 Scopus citations

Abstract

We developed a Liquid SCintillator (LSC) system for proton beam flux monitoring. A circulation system for a liquid scintillator is designed to the radiation damage to the LSC during proton beam irradiation. In this system, the LSC is made as thin as possible to minimize the energy loss because we want to use this detector as a trigger system for other detectors. The proton beam flux was measured using different beam currents at the MC-50 cyclotron at KIRAMS (Korea Institute of Radiological & Medical Sciences), and the results were compared. We can monitor the proton beam flux up to 106 proton/sec accurately by using a counting method with an ATMEGA128 embedded process equipped with a network system. Also, this system can be used to provide the trigger system for other detectors during the beam test.

Original languageEnglish
Pages (from-to)1506-1509
Number of pages4
JournalJournal of the Korean Physical Society
Volume50
Issue number5 PART I
DOIs
StatePublished - May 2007

Keywords

  • ATMEGA128
  • Beam flux monitoring
  • Liquid scintillator
  • MC-50 cyclotron

Fingerprint

Dive into the research topics of 'Development of liquid scintillator system for proton flux monitoring'. Together they form a unique fingerprint.

Cite this