TY - JOUR
T1 - Early failure of hemispheric hydroxyapatite-coated acetabular cups
AU - Kim, Shin Yoon
AU - Kim, Do Heon
AU - Kim, Yong Goo
AU - Oh, Chang Wug
AU - Ihn, Joo Chul
PY - 2006/5
Y1 - 2006/5
N2 - Total hip arthroplasties with hydroxyapatite coatings have shown encouraging results after early-term followup. We presumed hydroxyapatite- coating on a smooth hemispheric press-fit acetabular cup would enhance bone osseointegration and maintain stability of cup after midterm (minimum 5-year) followup. Sixty-three patients had 70 consecutive total hip arthroplasties. Five patients (eight hips) died from problems unrelated to surgery. The remaining patients (62 hips) were followed up for an average of 7 years (range, 6-9 years). The mean age of the patients was 49 years (range, 23-61 years). The average Harris hip score improved from 59 points (range, 32-82 points to 82 points (range, 37-100 points) at final followup. There were seven acetabular component revisions. Of the 55 unrevised cups, 47 hips (85%) were stable by bony ingrowth, five hips (9%) were fibrous stable, and three hips (5%) were unstable with cup migration. Osteolysis around the cup was observed in 18 hips (33%). The average polyethylene wear rate was 0.15 mm/year. Survival rates of the cups at 6 and 8 years were 94.3% and 60.5%, respectively. Total hip arthroplasties using an hydroxyapatite-coated smooth hemispheric acetabular cup showed an unexpected high failure rate in terms of fixation, occurrence of osteolysis, and revision after midterm followup.
AB - Total hip arthroplasties with hydroxyapatite coatings have shown encouraging results after early-term followup. We presumed hydroxyapatite- coating on a smooth hemispheric press-fit acetabular cup would enhance bone osseointegration and maintain stability of cup after midterm (minimum 5-year) followup. Sixty-three patients had 70 consecutive total hip arthroplasties. Five patients (eight hips) died from problems unrelated to surgery. The remaining patients (62 hips) were followed up for an average of 7 years (range, 6-9 years). The mean age of the patients was 49 years (range, 23-61 years). The average Harris hip score improved from 59 points (range, 32-82 points to 82 points (range, 37-100 points) at final followup. There were seven acetabular component revisions. Of the 55 unrevised cups, 47 hips (85%) were stable by bony ingrowth, five hips (9%) were fibrous stable, and three hips (5%) were unstable with cup migration. Osteolysis around the cup was observed in 18 hips (33%). The average polyethylene wear rate was 0.15 mm/year. Survival rates of the cups at 6 and 8 years were 94.3% and 60.5%, respectively. Total hip arthroplasties using an hydroxyapatite-coated smooth hemispheric acetabular cup showed an unexpected high failure rate in terms of fixation, occurrence of osteolysis, and revision after midterm followup.
UR - http://www.scopus.com/inward/record.url?scp=33744773455&partnerID=8YFLogxK
U2 - 10.1097/01.blo.0000195923.28480.d4
DO - 10.1097/01.blo.0000195923.28480.d4
M3 - Article
C2 - 16721953
AN - SCOPUS:33744773455
SN - 0009-921X
VL - 446
SP - 233
EP - 238
JO - Clinical Orthopaedics and Related Research
JF - Clinical Orthopaedics and Related Research