T1 - Effects of Exposure Parameters and Voxel Size for Cone-Beam Computed Tomography on the Image Matching Accuracy with an Optical Dental Scan Image
T2 - An in Vitro Study
AU - Mai, Hang Nga
AU - Lee, Du Hyeong
N1 - Publisher Copyright:
© 2021 Hang-Nga Mai and Du-Hyeong Lee.
PY - 2021
Y1 - 2021
N2 - This study is aimed at assessing the effects of exposure parameters and voxel size for cone-beam computed tomography (CBCT) on the image matching accuracy with an optical dental scan image. CBCT and optical scan images of a dry human mandible were obtained. Different CBCT settings were used: tube voltage, 60, 80, and 100 kVp; tube current, 6 and 8 mA; and voxel size, 100, 200, and 300 μm. Image matching between the CBCT and optical scan images was performed using implant planning software by dental professionals (n=18). The image matching accuracy in each combination of CBCT settings was evaluated by assessing the linear discrepancy between the three-dimensionally reconstructed radiological image and the registered optical scan image using an image analysis software program. The Kruskal-Wallis test and a post hoc Mann-Whitney U test with Bonferroni correction were used to compare the accuracy of image registration between the groups (α=0.05). Overall, the image matching accuracy was not significantly different between tube voltage and current settings; however, significantly higher image registration errors were found at the combination of 100 kVp tube voltage/8 mA tube current (F=8.44, P<0.001). Changes in voxel sizes did not significantly interfere with the image registration results. No interaction was found among voltage, current, and voxel size in terms of image registration accuracy (F=2.022, P=0.091). Different exposure parameter settings in tube voltage and tube current did not significantly influence the image matching accuracy between CBCT and optical dental scan images; however, a high radiation dose could be inappropriate. The image matching accuracy was not significantly affected by changing the voxel sizes of CBCT.
AB - This study is aimed at assessing the effects of exposure parameters and voxel size for cone-beam computed tomography (CBCT) on the image matching accuracy with an optical dental scan image. CBCT and optical scan images of a dry human mandible were obtained. Different CBCT settings were used: tube voltage, 60, 80, and 100 kVp; tube current, 6 and 8 mA; and voxel size, 100, 200, and 300 μm. Image matching between the CBCT and optical scan images was performed using implant planning software by dental professionals (n=18). The image matching accuracy in each combination of CBCT settings was evaluated by assessing the linear discrepancy between the three-dimensionally reconstructed radiological image and the registered optical scan image using an image analysis software program. The Kruskal-Wallis test and a post hoc Mann-Whitney U test with Bonferroni correction were used to compare the accuracy of image registration between the groups (α=0.05). Overall, the image matching accuracy was not significantly different between tube voltage and current settings; however, significantly higher image registration errors were found at the combination of 100 kVp tube voltage/8 mA tube current (F=8.44, P<0.001). Changes in voxel sizes did not significantly interfere with the image registration results. No interaction was found among voltage, current, and voxel size in terms of image registration accuracy (F=2.022, P=0.091). Different exposure parameter settings in tube voltage and tube current did not significantly influence the image matching accuracy between CBCT and optical dental scan images; however, a high radiation dose could be inappropriate. The image matching accuracy was not significantly affected by changing the voxel sizes of CBCT.
UR - http://www.scopus.com/inward/record.url?scp=85108564586&partnerID=8YFLogxK
U2 - 10.1155/2021/6971828
DO - 10.1155/2021/6971828
M3 - Article
C2 - 34222482
AN - SCOPUS:85108564586
SN - 2314-6133
VL - 2021
JO - BioMed Research International
JF - BioMed Research International
M1 - 6971828
ER -