Fracture resistances of heat-treated nickel-titanium files used for minimally invasive instrumentation

  • Yoo Jung Kang
  • , Sang Won Kwak
  • , Jung Hong Ha
  • , Gianluca Gambarini
  • , Hyeon Cheol Kim
  • Pusan National University
  • University of Rome La Sapienza

Research output: Contribution to journalArticlepeer-review

8 Scopus citations

Abstract

Background: This study compared the torsional resistance, bending stiffness, and cyclic fatigue resistances of different heat-treated NiTi files for minimally invasive instrumentation. Methods: TruNatomy (TN) and EndoRoad (ER) file systems were compared with ProTaper Gold (PG). Torsional load, distortion angle, and bending stiffness were assessed using a custom device AEndoS, and toughness was calculated using the torsional data. Cyclic fatigue resistance was evaluated using another custom device (EndoC) with 45-degree curved canal in which file was rotated until fracture using dynamic pecking motion at 37 °C. One-way analysis of variance and Duncan’s post-hoc comparison were conducted at a significance level of 95%. Scanning electron microscopy (SEM) analyzed fracture features and differential scanning calorimetry (DSC) analyzed phase transformation temperatures. Results: ER and TN showed significantly lower torsional strength than PG (p < 0.05). However, ER showed a significantly greater distortion angle and the highest toughness, followed by PG and TN (p < 0.05). Both ER and TN showed similar bending stiffness, which was lower bending stiffness than PG (p < 0.05). ER showed the highest cyclic fatigue resistance (p < 0.05). SEM revealed typical fracture features across all groups, with distinct milling grooves in PG and TN, not in ER. DSC indicated that PG and ER showed a peak of austenite (Ap) at temperatures higher than body temperature, 42 °C and 40 °C, respectively, while TN showed A p at 25 °C. Conclusion: Files for minimally invasive instrumentations typically exhibited high cyclic fatigue resistance but showed differences in the properties. The selection should depend on the root canal and tooth condition.

Original languageEnglish
Article number126
JournalBMC Oral Health
Volume25
Issue number1
DOIs
StatePublished - Dec 2025

Keywords

  • Bending stiffness
  • Cyclic fatigue
  • Heat treatment
  • Minimal invasive
  • Torsional resistance