Highly sensitive and novel point-of-care system, aQcare Chlamydia TRF kit for detecting Chlamydia trachomatis by using europium (Eu) (III) chelated nanoparticles

Ji Yeon Ham, Jaean Jung, Byung Gap Hwang, Won Jung Kim, Young Seop Kim, Eun Ju Kim, Mi Yeon Cho, Mi Sun Hwang, Dong Il Won, Jang Soo Suh

Research output: Contribution to journalArticlepeer-review

35 Scopus citations

Abstract

Background: The bacterium Chlamydia trachomatis is one of the leading causes of sexually transmitted diseases worldwide. Since no simple and effective tool exists to diagnose C. trachomatis infections, we evaluated a novel point-of-care (POC) test, aQcare Chlamydia TRF kit, which uses europium-chelated nanoparticles and a time-resolved fluorescence reader.

Methods: The test performance was evaluated by comparing the results obtained using the novel POC testing kit with those obtained using a nucleic acid amplification test (NAAT), using 114 NAAT-positive and 327 NAAT-negative samples.

Conclusions: This simple and novel test is a potential tool to screen a larger population, especially those in areas with limited resources.

Results: The cut-off value of the novel test was 20.8 with a detection limit of 0.27 ng/mL. No interference or cross-reactivity was observed. Diagnostic accuracy showed an overall sensitivity of 93.0% (106/114), specificity of 96.3% (315/327), positive predictive value (PPV) of 89.8% (106/118), and negative predictive value (NPV) of 97.5% (315/323). The sensitivity of the novel test was much higher than that of currently available POC tests. Furthermore, the relative ease and short turnaround time (30 min) of this assay enables C. trachomatis-infected individuals to be treated without a diagnostic delay.

Original languageEnglish
Pages (from-to)50-56
Number of pages7
JournalAnnals of Laboratory Medicine
Volume35
Issue number1
DOIs
StatePublished - 2015

Keywords

  • Chlamydia trachomatis
  • Europium
  • Point-of-care systems