Abstract
Since the 1980's, various types of implantable hearing aids using unique means for delivering acoustic power to the inner ear have been developed. Recently, implantable hearing aids that stimulate the round window by the middle ear transducer have received great attention because it reduces loading effect at the ossicular chain. In this study, we have implemented a direct install 3-pole type EM transducer in round window niche for implantable middle ear hearing aid. The 3- pole type EM transducer consists of two permanent magnets and three coils and exhibit structural features that minimize leakage flux, thereby permitting high efficiency and low magnetic field interference. The stapes velocity was measured using a laser Doppler vibrometer in response to the round window stimulation from the transducer. To verify the usefulness of the 3-pole type EM transducer, we compared the stapes vibration characteristics produced by the transducer and those from a sound source. The magnitude of stapes velocity due to the round window stimulation at 1 mArms was equivalent to that of stapes velocity at 94 dB SPL sound stimulation. Thus, the evaluation study shows that the 3-pole type EM transducer is suitable for implantable hearing devices.
| Original language | English |
|---|---|
| Pages (from-to) | 2503-2510 |
| Number of pages | 8 |
| Journal | Bio-Medical Materials and Engineering |
| Volume | 24 |
| Issue number | 6 |
| DOIs | |
| State | Published - 2014 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- 3-pole type EM transducer
- Implantable hearing aids
- Round window
- Stapes velocity
Fingerprint
Dive into the research topics of 'Implementation of a direct install 3-pole type em transducer in round window niche for implantable middle ear hearing aids'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver