In vitro study of HA and CMP grit-blasted and acid etched of Ti surface

A. W. Haryadi, C. K. You, S. Y. Kim, E. K. Park, S. Kim

Research output: Contribution to journalArticlepeer-review

Abstract

Grit-blasting using bioactive HA and biodegradable CMP followed by acid etching was done. The apatite formation of prepared Ti samples was evaluated by immersion in R-SBF. And cell viability, proliferation, and differentiation were conducted using MTT assay and ALP staining, In R-SBF immersion tests, non-etched HA-blasted samples showed the faster apatite-like formation than other samples. Acid etched and non-etched HA-blasted samples showed better cell viability and proliferation compared to CMP-blasted samples after 1 and 3 days. And the cell differentiation of non-etched HA-blasted samples was better compared to etched ones, and etched and non-etched CMP-blasted samples.

Original languageEnglish
Pages (from-to)395-398
Number of pages4
JournalKey Engineering Materials
Volume309-311 I
DOIs
StatePublished - 2006

Keywords

  • Acid Etching
  • Calcium Phosphate
  • Grit-blasting
  • In vitro

Fingerprint

Dive into the research topics of 'In vitro study of HA and CMP grit-blasted and acid etched of Ti surface'. Together they form a unique fingerprint.