T1 - Mechanical Properties of Various Glide Path Preparation Nickel-titanium Rotary Instruments
AU - Lee, Joo Yeong
AU - Kwak, Sang Won
AU - Ha, Jung Hong
AU - Abu-Tahun, Ibrahim H.
AU - Kim, Hyeon Cheol
N1 - Publisher Copyright:
© 2018 American Association of Endodontists
PY - 2019/2
Y1 - 2019/2
N2 - Introduction: The aim of this study was to compare the cyclic fatigue, torsional resistance, and bending stiffness of single-file glide path preparation nickel-titanium instruments. Methods: ProGlider (#16/progressive taper; Dentsply Sirona, Ballaigues, Switzerland), One G (#14/.03; Micro-Mega, Besancon, France), and EdgeGlidePath (#16/progressive taper; Edge Endo, Albuquerque, NM) were tested. For the bending stiffness test, the instrument was bent to a 45° angle along the main axis while 3 mm of the tip was secured. Cyclic fatigue resistance was measured during pecking movement, whereas the file was rotated within a reproducible simulated canal with a 3-mm radius and a 90° angle of curvature. The ultimate torsional strength and toughness were evaluated using a custom device. While tightly holding the end of the file at 3 mm, a rotational load of 2 rpm speed was applied until fracture. The results were analyzed using 1-way analysis of variance and Tukey post hoc comparison. A microscopic inspection was performed on the surface of a separate instrument using a scanning electron microscope. Results: EdgeGlidePath showed superior cyclic fatigue resistance compared with the other systems. One G showed higher maximum torsional strength than the others. One G showed the largest distortion angle and the highest toughness followed by EdgeGlidePath and ProGlider (P <.05). One G also showed larger bending stiffness than the others, whereas EdgeGlidePath showed a larger residual angle than the others (P <.05). Conclusions: Conventional wire showed higher toughness and torsional resistance than M-Wire and heat-treated nickel-titanium instruments. However, heat-treated wire showed higher cyclic resistance than conventional wire and M-Wire.
AB - Introduction: The aim of this study was to compare the cyclic fatigue, torsional resistance, and bending stiffness of single-file glide path preparation nickel-titanium instruments. Methods: ProGlider (#16/progressive taper; Dentsply Sirona, Ballaigues, Switzerland), One G (#14/.03; Micro-Mega, Besancon, France), and EdgeGlidePath (#16/progressive taper; Edge Endo, Albuquerque, NM) were tested. For the bending stiffness test, the instrument was bent to a 45° angle along the main axis while 3 mm of the tip was secured. Cyclic fatigue resistance was measured during pecking movement, whereas the file was rotated within a reproducible simulated canal with a 3-mm radius and a 90° angle of curvature. The ultimate torsional strength and toughness were evaluated using a custom device. While tightly holding the end of the file at 3 mm, a rotational load of 2 rpm speed was applied until fracture. The results were analyzed using 1-way analysis of variance and Tukey post hoc comparison. A microscopic inspection was performed on the surface of a separate instrument using a scanning electron microscope. Results: EdgeGlidePath showed superior cyclic fatigue resistance compared with the other systems. One G showed higher maximum torsional strength than the others. One G showed the largest distortion angle and the highest toughness followed by EdgeGlidePath and ProGlider (P <.05). One G also showed larger bending stiffness than the others, whereas EdgeGlidePath showed a larger residual angle than the others (P <.05). Conclusions: Conventional wire showed higher toughness and torsional resistance than M-Wire and heat-treated nickel-titanium instruments. However, heat-treated wire showed higher cyclic resistance than conventional wire and M-Wire.
KW - Cyclic fatigue resistance
KW - FireWire
KW - glide path file
KW - M-Wire
KW - nickel-titanium rotary file
KW - torsional fracture resistance
UR - http://www.scopus.com/inward/record.url?scp=85060719499&partnerID=8YFLogxK
U2 - 10.1016/j.joen.2018.10.017
DO - 10.1016/j.joen.2018.10.017
M3 - Article
C2 - 30711178
AN - SCOPUS:85060719499
SN - 0099-2399
VL - 45
SP - 199
EP - 204
JO - Journal of Endodontics
JF - Journal of Endodontics
IS - 2
ER -