TY - JOUR
T1 - Phantom evaluation of stacked-type dual-frequency 1-3 composite transducers
T2 - A feasibility study on intracavitary acoustic angiography
AU - Kim, Jinwook
AU - Li, Sibo
AU - Kasoji, Sandeep
AU - Dayton, Paul A.
AU - Jiang, Xiaoning
N1 - Publisher Copyright:
© 2015 Elsevier B.V. All rights reserved.
PY - 2015
Y1 - 2015
N2 - In this paper, we present phantom evaluation results of a stacked-type dual-frequency 1-3 piezoelectric composite transducer as a feasibility study for intracavitary acoustic angiography. Our previous design (6.5/30 MHz PMN-PT single crystal transducer) for intravascular contrast ultrasound imaging exhibited a contrast-to-tissue ratio (CTR) of 12 dB with a penetration depth of 2.5 mm. For improved penetration depth (>3 mm) and comparable contrast-to-tissue ratio (>12 dB), we evaluated a lower frequency 2/14 MHz PZT 1-3 composite transducer. Superharmonic imaging performance of this transducer and a detailed characterization of key parameters for acoustic angiography are presented. The 2/14 MHz arrangement demonstrated a -6 dB fractional bandwidth of 56.5% for the transmitter and 41.8% for the receiver, and produced sufficient peak-negative pressures (>1.5 MPa) at 2 MHz to induce a strong nonlinear harmonic response from microbubble contrast agents. In an in-vitro contrast ultrasound study using a tissue mimicking phantom and 200 μm cellulose microvessels, higher harmonic microbubble responses, from the 5th through the 7th harmonics, were detected with a signal-to-noise ratio of 16 dB. The microvessels were resolved in a two-dimensional image with a -6 dB axial resolution of 615 μm (5.5 times the wavelength of 14 MHz waves) and a contrast-to-tissue ratio of 16 dB. This feasibility study, including detailed explanation of phantom evaluation and characterization procedures for key parameters, will be useful for the development of future dual-frequency array transducers for intracavitary acoustic angiography.
AB - In this paper, we present phantom evaluation results of a stacked-type dual-frequency 1-3 piezoelectric composite transducer as a feasibility study for intracavitary acoustic angiography. Our previous design (6.5/30 MHz PMN-PT single crystal transducer) for intravascular contrast ultrasound imaging exhibited a contrast-to-tissue ratio (CTR) of 12 dB with a penetration depth of 2.5 mm. For improved penetration depth (>3 mm) and comparable contrast-to-tissue ratio (>12 dB), we evaluated a lower frequency 2/14 MHz PZT 1-3 composite transducer. Superharmonic imaging performance of this transducer and a detailed characterization of key parameters for acoustic angiography are presented. The 2/14 MHz arrangement demonstrated a -6 dB fractional bandwidth of 56.5% for the transmitter and 41.8% for the receiver, and produced sufficient peak-negative pressures (>1.5 MPa) at 2 MHz to induce a strong nonlinear harmonic response from microbubble contrast agents. In an in-vitro contrast ultrasound study using a tissue mimicking phantom and 200 μm cellulose microvessels, higher harmonic microbubble responses, from the 5th through the 7th harmonics, were detected with a signal-to-noise ratio of 16 dB. The microvessels were resolved in a two-dimensional image with a -6 dB axial resolution of 615 μm (5.5 times the wavelength of 14 MHz waves) and a contrast-to-tissue ratio of 16 dB. This feasibility study, including detailed explanation of phantom evaluation and characterization procedures for key parameters, will be useful for the development of future dual-frequency array transducers for intracavitary acoustic angiography.
KW - 1-3 composite transducer
KW - Dual-frequency transducer
KW - Intracavitary probe
KW - Superharmonic imaging
KW - Ultrasound contrast agents
UR - http://www.scopus.com/inward/record.url?scp=84947313822&partnerID=8YFLogxK
U2 - 10.1016/j.ultras.2015.06.009
DO - 10.1016/j.ultras.2015.06.009
M3 - Article
C2 - 26112426
AN - SCOPUS:84947313822
SN - 0041-624X
VL - 63
SP - 7
EP - 15
JO - Ultrasonics
JF - Ultrasonics
ER -