PTR, PCR and Energy Resolution Study of GAGG:Ce Scintillator

Pruittipol Limkitjaroenporn, Wiraporn Hongtong, Hong Joo Kim, Jakrapong Kaewkhao

Research output: Contribution to journalConference articlepeer-review

4 Scopus citations

Abstract

In this paper, the peak to total ratio (PTR), the peak to Compton ratio (PCR) and the energy resolution of cerium doped gadolinium aluminium gallium garnet (GAGG:Ce) scintillator are measured in the range of energy from 511 keV to 1332 keV using the radioactive source Na-22, Cs-137 and Co-60. The crystal is coupled with the PMT number R1306 and analyzed by the nuclear instrument module (NIM). The results found that the PTR and PCR of GAGG:Ce scintillator decrease with the increasing of energy. The results of energy resolution show the trend is decrease with the increasing of energy which corresponding to the higher energy resolution at higher energy. Moreover the energy resolution found to be linearly with.

Original languageEnglish
Article number012016
JournalJournal of Physics: Conference Series
Volume970
Issue number1
DOIs
StatePublished - 19 Mar 2018
Event2017 International Conference on Innovation in Education, Science and Culture, ICIESC 2017 - Medan, Indonesia
Duration: 8 Nov 20179 Nov 2017

Fingerprint

Dive into the research topics of 'PTR, PCR and Energy Resolution Study of GAGG:Ce Scintillator'. Together they form a unique fingerprint.