Skip to main navigation Skip to search Skip to main content

Vibration isolation system for cryogenic phonon-scintillation calorimeters

  • C. Lee
  • , H. S. Jo
  • , C. S. Kang
  • , G. B. Kim
  • , I. Kim
  • , S. R. Kim
  • , Y. H. Kim
  • , H. J. Lee
  • , J. H. So
  • , Y. S. Yoon
  • Institute for Basic Science
  • Korea Research Institute of Standards and Science

Research output: Contribution to journalArticlepeer-review

22 Scopus citations

Abstract

Cryogen-free dilution refrigerators are getting popular for rare event searches underground due to their advantages. However, the application of a pulse tube refrigerator introduces mechanical vibration that can translate into temperature fluctuation for calorimeters. The effect is significant in particular when the sensor is attached to a large absorber. A mechanical filter is installed to isolate the calorimeters from the vibration inside a cryogen-free dilution refrigerator while meeting thermal requirements.

Original languageEnglish
Article numberC02057
JournalJournal of Instrumentation
Volume12
Issue number2
DOIs
StatePublished - 16 Feb 2017

Keywords

  • Calorimeters
  • Cryocoolers
  • Cryogenics
  • Double-beta decay detectors

Fingerprint

Dive into the research topics of 'Vibration isolation system for cryogenic phonon-scintillation calorimeters'. Together they form a unique fingerprint.